||| 中文站 || German | Japanese | Russian | French | English |Testing Machine Forum ||
 
 
Current Location: Home » News » Technical Articles » 正文

ISO 1133 - Determination of the melt flow rate of thermoplastics

Increase font size  Decrease font size Date:2013-03-12  Views:2029
Core Tip:1 ScopeThis International Standard specifies two procedures for the determination of the melt mass-flow rate (MFR) and t

1 Scope

This International Standard specifies two procedures for the determination of the melt mass-flow rate (MFR) and the melt volume-flow rate (MVR) of thermoplastic materials under specified conditions of temperature and load. Procedure A is a mass-measurement method. Procedure B is a displacement-measurement method.
Normally, the test conditions for measurement of melt flow rate are specified in the material standard with a reference to this International Standard. The test conditions normally used for thermoplastics are listed in Annexes A and B.
The MVR will be found particularly useful when comparing materials of different filler content and when comparing filled with unfilled thermoplastics. The MFR can be determined from MVR measurements provided the melt density at the test temperature and pressure is known.
These methods are in principle also applicable to thermoplastics for which the rheological behaviour is affected during the measurement by phenomena such as hydrolysis, condensation or crosslinking, but only if the effect is limited in extent and only if the repeatability and reproducibility are within an acceptable range.
For materials which show significantly affected rheological behaviour during testing, these methods are not appropriate. In such cases, the use of the viscosity number in dilute solution, determined in accordance with the relevant part of ISO 1628, is recommended for characterization purposes.
NOTE The rates of shear in these methods are much smaller than those used under normal conditions of processing, and therefore data obtained by these methods for various thermoplastics may not always correlate with their behaviour during processing. Both methods are used primarily in quality control.
COORS melt flow indexer MFR400E is suitable for this test.
 
 
[ NewsSearch ]  [ ]  [ Tell a friend ]  [ Print this article ]  [ Close window ]

 

 
Recommended Graphic
Recommend News
Click Ranking
 
Contact QQ:1006931501*苏ICP备11049648号-15苏州亚诺天下仪器有限公司 版权所有
+86 18915433994
Buy pulling machine, preferred Coors brand! Contact Us | YANUO Group | Back Home【Service Hotline:+86 512 68559199】